Please use this identifier to cite or link to this item: https://idr.l2.nitk.ac.in/jspui/handle/123456789/7472
Title: A feasible QRS detection algorithm for arrhythmia diagnosis
Authors: Khadirnaikar, S.
Aparna, P.
Issue Date: 2017
Citation: 2016 International Conference on Advances in Electrical, Electronic and Systems Engineering, ICAEES 2016, 2017, Vol., , pp.32-37
Abstract: This paper presents a reliable QRS detection algorithm to detect and classify Electrocardiogram (ECG) waveform abnormalities by extracting features such as heart rate and duration of QRS complex. As R peak detection is the pivotal step in automatic electrocardiogram analysis, various mathematical operations like clipping, differentiation and squaring are carried out in the preprocessing stage to enhance the section containing the QRS complex. Thresholding is performed to detect the R peaks. In order to improve the accuracy of the algorithm search back technique is implemented to determine the missing R peaks and heart beat. Once the position of R peak is detected, positions of Q and S are determined using two different search intervals to take into account the anomalous conditions as well. The heart rate and QRS width are then computed and compared with the normal values to determine the degree and type of abnormality. MIT-BIH database is used to evaluate this algorithm. The algorithm gives sensitivity of 99.34% and positive predictivity of 96.79%. In the proposed algorithm complicated mathematical operations like Fourier Transform, Hilbert Transform or crosscorrelation are not computed, hence is convenient to realize. � 2016 IEEE.
URI: http://idr.nitk.ac.in/jspui/handle/123456789/7472
Appears in Collections:2. Conference Papers

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.