Please use this identifier to cite or link to this item: https://idr.l2.nitk.ac.in/jspui/handle/123456789/13629
Full metadata record
DC FieldValueLanguage
dc.contributor.authorVats, V.
dc.contributor.authorBaskaran, T.
dc.contributor.authorArya, S.B.
dc.date.accessioned2020-03-31T08:48:15Z-
dc.date.available2020-03-31T08:48:15Z-
dc.date.issued2018
dc.identifier.citationTribology International, 2018, Vol.119, , pp.659-666en_US
dc.identifier.uri10.1016/j.triboint.2017.11.023
dc.identifier.urihttp://idr.nitk.ac.in/jspui/handle/123456789/13629-
dc.description.abstractThe electrochemical corrosion and tribo-corrosion behaviors of nickel-free high nitrogen (HN SS) and high manganese containing austenitic stainless steel were studied in simulated body fluids such as Ringer's and artificial saliva solutions (ASS) using tribo-meter attached with the potentiostat. Type 316L SS used as reference alloy for comparison. Open circuit potential (OCP) and potentiodynamic polarization techniques were used to examine the passivation and corrosion behavior of both the stainless steels under the applied loads of 5 and 10 N at room temperature and also compared with the static condition of corrosion. Pitting resistance of HN SS was found to be significantly higher over type 316L SS. � 2017 Elsevier Ltden_US
dc.titleTribo-corrosion study of nickel-free, high nitrogen and high manganese austenitic stainless steelen_US
dc.typeArticleen_US
Appears in Collections:1. Journal Articles

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.